Page 29 - Energize July 2022
P. 29

NEWS




           Hydraulic deadweight tester for


           calibrating pressure measuring


           instruments





                eadweight testers are accepted   there is a balance of forces between pressure and wheel weights. Due to the
                as the most accurate method for   excellent quality of the system this pressure remains stable over several minutes, so
          Dthe calibration of electronic or    that, for instance, adjustments of the device under test can be carried out without
           mechanical pressure measuring devices.   any problems.
           Instrotech is offering the LR-Cal LDW series   For users who require a digital pressure display for connection to the
           of deadweight testers, which are available   deadweight tester, Instrotech can supply the Keller Leo2 that offers accuracies of
           in the ranges from 10 bar to 1400 bar.  0,1%. An intrinsically safe version is available.
             The direct measurement of pressure
           (according to its definition as a quotient   Contact Instrotech, Phone 010 595-1831, sales@instrotech.co.za
           of force and area) ensures high accuracies   Facebook: bit.ly/3rj1SWY
           and long-term stability. These testers are   LinkedIn: bit.ly/3rkfKjL
           used extensively in calibration laboratories   YouTube: bit.ly/3irItiy
           in industry, standards institutes and   Product: https://bit.ly/3zNwYKH
           research facilities.                Hastags: #pressurecalibration #calibration #Instrotech
             The LDW series are ideally suited
           to on-site use as well as for service and
           maintenance purposes. The customized
           assembly allows set up of a complete,
           compact system consisting of a universal
           basement and the measuring systems
           (piston/cylinder units). The high-quality
           piston/cylinder units are manufactured
           from tungsten carbide which is known
           for its small temperature co-efficiency.
           Fast and uncomplicated changes of the
           measuring range are possible using a flat
           spanner.
             In order to generate the individual
           test points, the piston/cylinder system
           is weighted with mass loads which are
           also calibrated and specially adapted to
           the required application. The basement
           is fitted with the corresponding system,
           depending on the measuring range of
           the device under test. The pressure is set
           via an integrated pump or, if an external
           pressure supply is available, via control
           valves (pneumatic models).
             For fine adjustment, an adjustable
           volume with precision spindle is available.
           The weight applied is proportional to the
           desired pressure and provided by using
           optimally graduated weights. As soon as
           the measuring system reaches equilibrium,





                                                    energize | July 2022 | 27