Page 35 - EngineerIt August 2021
P. 35

MEASUREMENT



        Where: EHbO 2,red = extinction coefficient of HbO 2 at 600 nm,
        E HbO2,ired = extinction coefficient of HbO 2 at 940 nm

        E RHb,ired = extinction coefficient of RHb at 940 nm, E RHb,red =
        extinction coefficient of RHb at 600 nm

        However, the Beer-Lambert law cannot be used directly as
        there are a number of variable factors in every optical design
        that cause variations to the RoR to SpO 2 relationship. These
        include mechanical baffle design, LED to PD spacing, electronic
        and mechanical ambient light rejection, PD gain errors, and
        many more.
           To obtain clinical grade accuracy from a PPG-based SpO 2
        pulse oximeter, a lookup table or algorithm must be developed for
        the correlation between RoR and SpO 2.

        Calibration                                            Figure 4: LED-PD configuration
        Calibration of the measurement system is required to develop
        a high accuracy SpO 2 algorithm. To calibrate an SpO 2 system,   Reflective PPG configurations are chosen when the PD and
        a study must be completed where a participant’s blood oxygen   LED must be placed next to each other for practicality, such as
        levels are medically reduced, monitored and overseen by a   with wrist- or chest-worn devices.
        medical professional. This is known as a hypoxia study.
           The SpO 2 measurement system can only be as accurate as   Sensor positioning and perfusion index
        the reference. Reference options include medical grade finger   Positioning on the wrist and chest require greater dynamic range
        clip pulse oximeters and the gold standard co-oximeter. The   in the PPG AFE as the DC signal is greatly increased due to the
        co-oximeter is an invasive method of measuring the oxygen   depth of the arteries below static reflective components such as
        saturation of blood that yields high accuracy, but in most cases is   skin, fat and bone.
        not convenient to administer.                             Greater resolution in the PPG measurements will reduce the
           The calibration process is used to generate a best fit curve   uncertainty in the SpO 2 algorithm. With a typical PI of 1% to 2%
        of RoR value calculated from the optical SpO 2 device to the co-  for wrist-worn SpO 2 sensors, the goal of pulse oximeter design is
        oximeter SaO 2 measurement. This curve is used to generate a   to increase the PI through mechanical design or to increase the
        lookup table or equation for calculating SpO 2.        dynamic range.
           Calibration will be required for all SpO 2 designs as RoR is   The spacing of the LED to PD will have a major effect on
        dependent on a number of variables such as LED wavelength   the PI. Too little spacing will increase LED to PD crosstalk or
        and intensity, PD response, body placement and ambient light   backscatter. This will appear as a DC signal and saturate the AFE.
        rejection, which will differ with each design.            Increasing this spacing reduces the effect of both backscatter
           An increased perfusion index and, in turn, a high AC dynamic   and crosstalk but also reduces the current transformer ratio
        range on the red and IR wavelengths will increase the sensitivity   (CTR), which is the LED output to PD return current. This will
        of the RoR calculation and, in turn, return a more accurate SpO 2   affect the efficiency of the PPG system and require greater LED
        measurement.                                           power to maximise the AFE dynamic range.
           During a hypoxia study, 200 measurements equally spaced   Rapidly pulsing one or multiple LEDs has the benefit of
        between 100% and 70% blood oxygen saturation need to be   reducing the 1/f noise contribution to the overall signal. Pulsing
        recorded. Subjects are chosen with a variety of coloured skin   the LEDs also makes it possible to use synchronised modulation
        tone and an equal spread of age and gender. This variation in
        skin tone, age and gender accounts for differing perfusion index
        results from a spread of individuals.
           The overall error for transmissive pulse oximeters must be
        ≤3.0% and ≤3.5% for reflective configuration.
        DESIGN CONSIDERATIONS
        Transmissive vs. reflective
        A PPG signal can be obtained using a transmissive or reflective
        LED and PD configuration. A transmissive configuration measures
        the non-absorbed light passed through a part of the body. This
        configuration is best suited to areas such as the finger and
        earlobe where measurement benefits from the capillary density
        of these body locations, which make the measurements more
        stable, repetitive and less sensitive to variations in placement.
        Transmissive configurations achieve a 40 dB to 60 dB increase in
        the perfusion index.                                   Figure 5: ADI VSM watch V4, baffle, and LED DP array



                                                   EngineerIT | August 2021 | 33
   30   31   32   33   34   35   36   37   38   39   40